Implement Industry Ideas
  • PH analyzer / TDS / conductivity / salinity

C6010

Article: 00105924

PH analyzer / TDS / conductivity / salinity

pH - mV - Conductivity - Resistivity - Salinity - TDS - Dissolved oxygen - Air pressure - Temperature

one common input for all measurements

Specifications

pH: -2...+16 pH
mV: ±2000 mV
Conductivity: 0...2000 mS/m
Resistivity: 0...200 M?.cm
Salinity: 0.0...70.0
TDS: 0...100 g/l
Dissolved oxygen: 0...60mg/l 0...600%
Air pressure: 600...1300hPa
Temperature: -5...+105°C

pH

Mulit-point (1...5) calibration for more linearity.
Selectable reslution from 0.001 pH to 0.1 pH.
Automatic calibration with any of eleven pre-programmed and five user specified pH buffers.
Create your own buffer/temperature tables!
Accepts mV electrodes with any zero point(Eo) between ±999mV.

mV

Features mV calibration for accurate ORP measurements
Selectable resolution from 0.1 mV to 1 mV

Ion

Direct concentration measurement.
Multi-point (2...5) calibration and an additional blank correction for measuring low concentrations.

Conductivity

Multi-point (1...3) calibration for more linearity.
An electrode with a typical cell constant of 1 cm-1(standard) permits to measure from 0.01 µS/cm to 200 mS/cm in five ranges.
An electrode with a typical cell constant of 0.1 cm-1 permits to measure from 0.001µS/cm to 20 mS/cm in five ranges.
An electrode with a typical cell constant of 10 cm-1 permits to measure from 0.1µS/cm to 2000 mS/cm in five ranges.
Automaticly selects correct range and frequency.
Selectable reference temperature: 20°C or 25°C

Specifications

  • Measured value: pH, TDS, conductivity, salinity, resistivity
  • Gas type: dissolved oxygen
  • Portability: portable
  • Other characteristics: temperature

Other items

In this category | In this category, (CONSORT)

Leave request for price calculation right now, we will send you best prices.