Article: 00083215

Multi-parameter calibration and test bench / for dosimeter

The Calibrator/M consists of:

A dosimeter handling and positioning conveyor drawer:
A test slot allows positioning the dosimeter for exposure to the 137Cs control source. A 137Cs source and its lead shield, designed as a collimator:
Activity of the source of 137Cs - 1.11 MBq (30 µCi). The source is placed 3.5 cm (1.4 in.) from the DOSICARD's detector. Dose rate - 88 µSv/h (8.8 mrem/hr). Precision of the measurement - ±6%. Irradiation time of the dosimeter - from four to 15 minutes (specified by the operator).
Note: The dose rate at the lead shield is

Specifications

  • Test type: multi-parameter
  • Test material: for dosimeter

Other items

In this category | In this category, (Canberra Industries)

Leave request for price calculation right now, we will send you best prices.